Cut the sample from the plant-
Preparing a sample for the pressure chamber usually involves removing
a small lateral branch; if very small plants are used, the entire tip may
be required. Removing the sample with a sharp knife or razor blade
eliminates needless re-cutting, which can introduce considerable error
in some species. The cut should be clean and normally at an angle
makes for easier viewing.

Seal the sample into the chamber lid-
Next you will need to seal the sample into the chamber this can be
done by using a Compression Gland
Cover or by using gaskets, inserts
& insertion tools. Click on the one that applies to you and
then return to this page to finish the tutorial on "Making the Measurement".
An erroneous measurement may result if a large proportion of a small, succulent sample (e.g. conifer needle, grass blade, new shoot, etc.) is squeezed by the gasket. For example, PMS can be successfully measured with a fascicle of pine needles, but substantial (5 to 10 bar) errors can occur in measuring needles from Douglas fir, even with a very thin gasket. So long as no more than 10% of the sample protrudes through or is squeezed by the gasket, this type of error should be negligible.
Making the measurement-
The control valve is turned to CHAMBER to increase the pressure inside
the chamber. The cut surface of the plant sample which protrudes
through the chamber cover is observed carefully. When water appears
at the cut surface the control valve is turned to OFF immediately, and
the pressure indicated on the gauge is recorded. The control valve
is then turned to EXHAUST and when the chamber pressure reaches zero (indicated
by the gauge), the chamber cover and sample can be removed. The instrument
is ready for the next measurement.


When the endpoint is determined and nitrogen turned off with the control valve, there is a slight decrease in pressure even if there is no leak. This phenomenon is particularly noticeable at high stress levels with a rapid rate of pressure increase. It is caused by temperature changes inside the chamber and can be minimized by adjusting the rate of pressure increase.
Resin can obscure the endpoint and may be bothersome in certain species, particularly with pines. Resin forms bubbles which break and can be wiped away to allow observation of the true endpoint. Resin usually appears at very low pressures and need not interfere with accurate measurements. With experience, resin can be dealt with satisfactorily.
CAUTION! Never place any part of your body directly over the chamber when the chamber is pressurized. It is possible that the sample may be forced out of the chamber at high velocity!